|
Surface Chemistry |
|||||
| NRL / Materials / Chemistry / Code 6170 / Code 6177 / Research Areas / AFM: General Concept | NRL Resources | ||||
|
Atomic Force Microscopy: General Concept and Defining Characteristics Scanned-proximity probe microscopes provide very high resolution images of various sample properties The atomic force microscope is one of about two dozen types of scanned-proximity probe microscopes. All of these microscopes work by measuring a local property - such as height, optical absorption, or magnetism - with a probe or "tip" placed very close to the sample. The small probe-sample separation (on the order of the instrument's resolution) makes it possible to take measurements over a small area. To acquire an image the microscope raster-scans the probe over the sample while measuring the local property in question. The resulting image resembles an image on a television screen in that both consist of many rows or lines of information placed one above the other. Unlike traditional microscopes, scanned-probe systems do not use lenses, so the size of the probe rather than diffraction effects generally limit their resolution. The atomic force microscope measures topography with a force probe
AFM (figure 1) operates by measuring attractive or repulsive forces between a tip and the sample (Binnig et al., 1986). In its repulsive "contact" mode, the instrument lightly touches a tip at the end of a leaf spring or "cantilever" to the sample. As a raster-scan drags the tip over the sample, some sort of detection apparatus measures the vertical deflection of the cantilever, which indicates the local sample height. Thus, in contact mode the AFM measures hard-sphere repulsion forces between the tip and sample. In noncontact mode, the AFM derives topographic images from measurements of attractive forces; the tip does not touch the sample (Albrecht et al., 1991). Because it does not allow the imaging of samples under water, I have not used the attractive mode. AFMs can achieve a resolution of 10 pm, and unlike electron microscopes, can image samples in air and under liquids. In principle, AFM resembles
the record player as well as the stylus profilometer. However, AFM incorporates
a number of refinements that enable it to achieve atomic-scale resolution:
Laser beam deflection offers a convenient and sensitive method of measuring cantilever deflection AFMs can generally measure the vertical deflection of the cantilever with picometer resolution. To achieve this most AFMs today use the optical lever, a device that achieves resolution comparable to an interferometer while remaining inexpensive and easy to use (Meyer et al., 1988; Alexander et al., 1989). The optical lever (figure 1) operates by reflecting a laser beam off the cantilever. Angular deflection of the cantilever causes a twofold larger angular deflection of the laser beam. The reflected laser beam strikes a position-sensitive photodetector consisting of two side-by-side photodiodes. The difference between the two photodiode signals indicates the position of the laser spot on the detector and thus the angular deflection of the cantilever. Because the cantilever-to-detector distance generally measures thousands of times the length of the cantilever, the optical lever greatly magnifies motions of the tip. Because of this ~2000-fold magnification optical lever detection can theoretically obtain a noise level of 10-14 m/Hz1/2 Putman et al., 1992). For measuring cantilever deflection, to date only the relatively cumbersome techniques of interferometry and tunneling detection have approached this value. AFM cantilevers have high flexibility
A high flexibility stylus exerts lower downward forces on the sample, resulting in less distortion and damage while scanning. For this reason AFM cantilevers generally have spring constants of about 0.1 N/m (figure 2). As Paul Hansma points out, a Slinky is about ten times stiffer (1 N/m). It would take a very long time
to image a surface by dragging a Slinky over it (in the configuration
of figure 2), because a Slinky cannot respond quickly as it passes over
features. That is, a Slinky has a low resonant frequency, but an AFM cantilever
should have a high resonant frequency.
Micromachining techniques produce inexpensive, reasonably sharp tips
A close enough inspection of any AFM tip reveals that it is rounded off. Therefore force microscopists generally evaluate tips by determining their "end radius." In combination with tip-sample interaction effects, this end radius generally limits the resolution of AFM. As such, the development of sharper tips is currently a major concern. Force microscopists generally use one of three types of tip. The "normal tip" (figure 4a; Albrecht et al., 1990) is a 3 µm tall pyramid with ~30 nm end radius. The electron-beam-deposited (EBD) tip or "supertip" (figure 4b; Keller and Chih-Chung, 1992) improves on this with an electron-beam-induced deposit of carbonaceous material made by pointing a normal tip straight into the electron beam of a scanning electron microscope. Especially if the user first contaminates the cantilever with paraffin oil, a supertip will form upon stopping the raster of the electron beam at the apex of the tip for several minutes. The supertip offers a higher aspect ratio (it is long and thin, good for probing pits and crevices) and sometimes a better end radius than the normal tip. Finally, Park Scientific Instruments offers the "Ultralever" (figure 4c), based on an improved microlithography process. Ultralevers offers a moderately high aspect ratio and on occasion a ~10 nm end radius. Tube piezoceramics position the tip or sample with high resolution
Piezoelectric ceramics are a class of materials that expand or contract when in the presence of a voltage gradient or, conversely, create a voltage gradient when forced to expand or contract (Gallego-Juárez, 1989). Piezoceramics make it possible to create three-dimensional positioning devices of arbitrarily high precision. Most scanned-probe microscopes use tube-shaped piezoceramics because they combine a simple one-piece construction with high stability and large scan range. Four electrodes cover the outer surface of the tube, while a single electrode covers the inner surface. Application of voltages to one or more of the electrodes causes the tube to bend or stretch, moving the sample in three dimensions (figure 5). AFMs use feedback to regulate the force on the sample
The presence of a feedback loop is one of the subtler differences between AFMs and older stylus-based instruments such as record players and stylus profilometers. The AFM not only measures the force on the sample but also regulates it, allowing acquisition of images at very low forces. The feedback loop (figure 6) consists of the tube scanner that controls the height of the entire sample; the cantilever and optical lever, which measures the local height of the sample; and a feedback circuit that attempts to keep the cantilever deflection constant by adjusting the voltage applied to the scanner. One point of interest: the faster the feedback loop can correct deviations of the cantilever deflection, the faster the AFM can acquire images; therefore, a well-constructed feedback loop is essential to microscope performance. AFM feedback loops tend to have a bandwidth of about 10 kHz, resulting in image acquisition times of about one minute. |
||||||||||||||||||||||||||||
| Privacy Policy |
Code 6170 |
|